What type of input signals does this module support?
The Mitsubishi FX2N-4AD-TC supports thermocouple input signals, specifically Type K and Type J thermocouples.
How many input channels are available on this module?
This module provides four input channels for thermocouple connections.
What is the power supply requirement for this module?
The module requires a 24V DC power supply with a tolerance of ±10%.
What is the resolution of the temperature readings?
The resolution for Type K thermocouples is 0.4°C (0.72°F) and for Type J thermocouples is 0.3°C (0.54°F).
Can this module be used with other Mitsubishi PLC series?
Yes, the FX2N-4AD-TC can be connected to FX0N, FX1N, FX2N, FX2NC, FX3G, FX3U, and FX3UC series PLCs.
What are the dimensions of this module?
The dimensions of the module are 55 mm in width, 87 mm in height, and 90 mm in depth.
What is the weight of the module?
The module weighs approximately 0.3 kg (0.66 lbs).
What is the conversion speed of the module?
The conversion speed is 240ms ± 2% per channel.
What is the overall accuracy of the temperature measurements?
The overall accuracy is ±0.5% of full scale plus 1°C.
How many I/O points does this module occupy on the PLC expansion bus?
The FX2N-4AD-TC occupies 8 I/O points on the PLC expansion bus.
What type of isolation is provided between the analog and digital circuits?
The module features photo-coupler isolation between the analog and digital circuits.
What is the operating temperature range for Type K thermocouples?
For Type K thermocouples, the operating temperature range is -100°C to +1,200°C (-148°F to +2,192°F).
What is the operating temperature range for Type J thermocouples?
For Type J thermocouples, the operating temperature range is -100°C to +600°C (-148°F to +1,112°F).
What should be done if a thermocouple is disconnected?
If a thermocouple is disconnected, the module will latch the error status, and the temperature data before the error will be retained. The error can be cleared by writing K0 to BFM #28 using the TO instruction or by turning off the power.
What is the recommended length for compensating cables to avoid noise interference?
It is recommended to use compensating cables shorter than 100 meters to avoid noise interference.